Model Number RD900
Device Problem Device Handling Problem (3265)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) method: the complaint rd900 neopuff infant resuscitator was returned to the f&p service center in (b)(4) where it was inspected by a trained f&p service technician.Our investigation is based on the service report and photograph provided by the f&p service technician.Results: the photograph provided revealed that the gas outlet port of the subject neopuff was broken.Conclusion: the neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to an infant until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to damage, for instance when accidentally dropped or subjected to considerable external force.The physical damage observed was most likely due to impact to the subject neopuff unit.The neopuff technical manual states the following: "dropping the neopuff infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient." each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff unit was repaired and returned to the hospital after passing the performance tests specified on the neopuff technical manual.
 
Event Description
A hospital in (b)(6) reported that the gas outlet port of a neopuff infant resuscitator was broken.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Type of Device
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key7649217
MDR Text Key112885407
Report Number9611451-2018-00556
Device Sequence Number1
Product Code BTL
UDI-Device Identifier09420012410931
UDI-Public010942001241093110210014785311170207
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number2100147853
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-