MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/31/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump had a flashing display.Troubleshooting was performed.Blank display re-accrued.Device is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
Device had blank flashing white lcd due to cracked lcd controller and cracked or bleeding lcd glass.Unable to verify missing segments due to blank flashing white lcd.Unable to perform the self test, sleep current measurement, active current measurement and displacement test due to blank flashing white lcd.Device received without the original battery cap.Unable to verify damage to the original battery cap.
|
|