(b)(4).Device evaluation by manufacturer: field service engineering (fse) contacted the customer over-the-phone to address the reported issue.During troubleshooting, fse instructed the customer to look for leaks and the customer found a leak at the filter housing.Fse then instructed the customer to change the filter and remove the leak from the system.The customer adjusted the flow rate down to adjust the sa1c retention time (rt) to 0.59 minutes.The instrument was verified as operational.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were two similar complaints found during the searched period.The g8 operator's manual under chapter 1, introduction and applications, states the following: the time from injection of the sample to the time the specific peak elutes off the column is called retention time.The tosoh automated glycohemoglobin analyzer hlc-723g8 software has been written so that each of the expected fractions has a window of acceptable retention times.If the designated peak falls within the expected window, the chromatogram peaks will be properly identified.When a peak elutes at a retention time not within a specified window, an unknown peak (p00) results.If more than one peak elutes at times not specified by the software windows, each is given a sequential p0x title.In order to keep the peaks within their appropriate windows, it may be necessary to change how fast or slow the buffers are moving through the system by changing the pump flow rate.The g8 variant analysis mode training manual under lesson 8 - troubleshooting, states the following: adjusting the flow rate - how and why on the tosoh automated glycohemoglobin analyzer hlc-723g8; variant analysis mode it may be necessary to adjust the flow rate because either unidentifiable peaks appear on all the chromatograms or the average retention time for various peaks has changed significantly.The flow rate is changed by changing the flow factor in the instrument.The flow factor is generally 1.00 ml/min.The flow factor should only be adjusted +/- 0.05 of the default factory setting.Chapter 5 maintenance procedure under section 5.7 filter replacement states the following: press the pump key again to start the elution buffer delivery.Make sure that the pressure falls within a range less than the original pressure (which is indicated on the column inspection report) + 4 mpa and that there are no leaks from the filter housing components and tube connections.If a leak is found, tighten the assembly further.The most probable cause of the high retention times was due to a leak at the filter housing.
|