WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; LRC INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number C-CAE-14.0-70-FIC
Device Problem Material Frayed (1262)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).This complaint is logged to capture the additional incidents referred to in the customer report to ansm "this incident happened several times".Name and address for importer site: (b)(4).Similar to device under 510(k) e597079.Investgation is still in progress.
 
Event Description
Description of event according to initial reporter."difficult intubation required the use of type cook eschmann tube, without difficulty during use.A fibroscopic check immediately after the intubation helped to display an endobronchial foreign matter, which according to the department reporter, would be ¿peeling¿ of the tube.Please note that the probe used was a carlens shiley 39f probe, from covidien.This incident happened several times".Patient outcome: "clinical consequences reported are moderate.Increase of intubation period".
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Exemption number: e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Manufacturer ref#: (b)(4).Mfr site: name and address for importer site: (b)(4).Ec method code: 4109 - historical data analysis 4114 - device not returned.Summary of investigational findings: no product was returned to assist the investigation, but it is reported that the peeling was noticed during placement of a double lumen tube and the frova introducer is designed for placement of a single lumen tube only.However, it is noted that the area representative is focusing on the fact that "the problems were always with dlt".Ifu, intended use: "the 14.0 french catheter introducer has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.Note: do not use the frova intubating introducer with double lumen endotracheal or endobronchial tubes." also, under warnings is stated "do not use the frova intubating introducer with double lumen endotracheal or endobronchial tubes." no evidence to suggest that the device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FROVA INTUBATING INTRODUCER
Type of Device
LRC INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key7649491
MDR Text Key112706495
Report Number3002808486-2018-00789
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC-CAE-14.0-70-FIC
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Distributor Facility Aware Date06/26/2018
Date Manufacturer Received07/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-