• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2426-0500
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2018
Event Type  malfunction  
Manufacturer Narrative
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
 
Event Description
The customer reported noticing a hole 3 inches above the distal injection port of the tubing while infusing ns to a patient.The infusion was stopped while a replacement set was obtained.There was no patient harm.
 
Event Description
The customer reported noticing a hole 3 inches above the distal injection port of the tubing while infusing ns to a patient.The infusion was stopped while a replacement set was obtained.There was no patient harm.
 
Manufacturer Narrative
The customer¿s report of a hole in the tubing was confirmed.Visual inspection showed that the pvc tubing had a slice cut 3 inches above the distal smartsite inlet port.Functional and pressure testing confirmed leaking from the cut.The cause of the hole in the tubing was identified as a slice cut in the tubing consistent with damage from a sharp object.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7651420
MDR Text Key112764373
Report Number9616066-2018-00954
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2426-0500
Device Catalogue Number2426-0500
Other Device ID Number7613203020992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Date Manufacturer Received06/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age72 YR
Patient Weight71
-
-