• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE FAIRFIELD BEQ-TOP 49101 CHILDRENS LARGE ECC PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD BEQ-TOP 49101 CHILDRENS LARGE ECC PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 701067069R01
Device Problems Break (1069); Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
No decontamination or visual inspection of product was performed since product was not returned.A photograph was provided by the customer and it was evaluated.The photo indicated that a stopcock was disconnected from the pressure line, a pigtail was disconnected, and a pressure line was broken off.The reported event of loose and broken components was confirmed.The most likely cause could not be determined.A lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.(b)(4).
 
Event Description
It was reported by the customer that custom tubing pack packaging has been getting progressively worse.The customer found that there were numerous loose connections, undone connections, and broken stopcocks.Also the customer stated that the pigtail was totally broken, a stopcock off of its pigtail, and a pigtail not even on the line.While priming one pigtail was broken and needed to be replaced.The customer had been informed that on closed caps on the stopcock, once you loosen them to de-air the port, the caps do not lock back down always.Many just spring and don't fully tighten.There was no patient involvement and therefore no reported injury.The date of the event is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEQ-TOP 49101 CHILDRENS LARGE ECC PACK
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key7652308
MDR Text Key113020309
Report Number2248146-2018-00410
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K08059223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/18/2019
Device Catalogue Number701067069R01
Device Lot Number3000059724
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age YR
Date Manufacturer Received06/05/2018
Date Device Manufactured09/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-