Model Number CI-1400-01
Device Problem Extrusion (2934)
Patient Problems Failure of Implant (1924); Unspecified Infection (1930); Inflammation (1932); Skin Tears (2516)
Event Type  Injury  
Event Description
The recipient reportedly experienced electrode extrusion and an infection.The recipient presented with broken skin and inflammation around the implant site.The recipient's device was explanted.The recipient was not reimplanted.
 
Manufacturer Narrative
Additional information regarding treatment details will not be provided.The recipient's infection reportedly resolved.The external visual inspection revealed the electrode was severed near the electrode ground ring and near the array prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key7652443
MDR Text Key112799556
Report Number3006556115-2018-00298
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup
Report Date 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/30/2007
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-