Device is a combination product.(b)(4).Device evaluated by mfr.: promus premier ous mr 3.50 x 38mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified no issues.The stent showed no signs of damage or strain and was set equidistant between the proximal and distal markerbands.The crimped stent od (outer diameter) was measured and was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon had not been subjected to any significant positive pressure.A visual and tactile examination of the device revealed a hypotube break at 260mm distal to the distal end of the strain relief.There were multiple hypotube kinks noted.An examination of the inner and outer shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage or strain.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Reportable based on device analysis completed on 08-jun-2018.It was reported that shaft kink occurred.Vascular access was obtained via the femoral artery.The 90% stenosed, concentric target lesion containing a significant bend of >=90 degrees was located in the mildy calcified mid left anterior descending artery.A 38 x 3.50mm promus premier¿ drug-eluting stent was advanced to treat the lesion.However, it was noticed that stent shaft got an acute kink.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a hypotube break.
|