• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION ZOLL IVTM THERMOGARD XP; THERMAL REGULATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION ZOLL IVTM THERMOGARD XP; THERMAL REGULATING SYSTEM Back to Search Results
Model Number 8700-0650
Device Problems Device Issue (2379); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2018
Event Type  malfunction  
Manufacturer Narrative
The report of the thermogard console sn (b)(4) displaying "factory configuration not complete" error message upon power up was confirmed during functional testing and in the event log.Upon visual inspection the side panel of the console was dented and front housing cover is damage.These observation are not related to the reported event.Evaluation of the console revealed "factory configuration not complete" error message during initial power up.The cause of the error message was attributed to a defective input/output printed circuit board.The analysis of the event log revealed multiple mid27-3 (read cycle timeout) error messages and one mid27-4 (eeprom crc failure) on the reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for the thermogard console with serial number (b)(4).
 
Event Description
During setup of the themogard console sn (b)(4) displayed " factory configuration not complete error" error message on the display panel screen upon all power-ups.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM THERMOGARD XP
Type of Device
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7652993
MDR Text Key112858640
Report Number3010617000-2018-00695
Device Sequence Number1
Product Code NCX
UDI-Device Identifier00849111075022
UDI-Public00849111075022
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072234
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0650
Device Catalogue Number8700-0650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-