Catalog Number C6855
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
Inspection of the device found that the device is not working properly and the tip was found to be broken.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: cautions: inspect instrument prior to use to ensure proper function; no loose pins or misalignment, and that the instrument is in good physical condition.Do not use excessive force on instruments to avoid damage or breakage during use.Avoid unintended contact with other surgical instruments during use to prevent damage or breakage.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event Description
This distributor reported that the c6855, concept linear wide bite punch, straight, device tip broke during an acl reconstruction (with meniscectomy) procedure on (b)(6) 2018.There was no delay or cancellation to the procedure.It is reported that no part of the device fell into the surgical site; therefore, there were no components to retrieve.The reporter has stated that there was no harm to the user or the patient due to this event.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCEPT LINEAR WIDE BITE PUNCH, STRAIGHT
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
TECOMET, INC.
5307 95th ave.
kenosha WI 53144
Manufacturer (Section G)
TECOMET, INC.
5307 95th ave.
kenosha WI 53144
Manufacturer Contact
john berga
11311 concept blvd.
largo, FL 33773
7273995358
MDR Report Key7654797
MDR Text Key113017485
Report Number1017294-2018-00101
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeBG
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberC6855
Device Lot Number0118C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2018
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Age48 YR
Patient Weight72
-
-