Model Number C12059
Device Problems Fracture (1260); Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
The reported event of guidewire fracture was confirmed.The results of the investigation concluded that the shaft had been fractured and separated, with subsequent fracture of the microcables; the corewire remained intact and exposed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the guidewire damage is consistent with damage during use.The pressurewire x instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire x instructions for use states that the pressurewire is a delicate instrument and should be handled carefully.Bending or excessive force during removal from the packaging coil may damage the guidewire.The pressurewire x instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.
 
Event Description
Upon insertion to guide, the pressurewire x, wireless snapped in half.Both halves were retrieved without issue since the device had not yet entered the patient.The procedure was completed with another pressurewire x, wireless.No patient weight or age was released from the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ X, WIRELESS, RX 175CM
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7655292
MDR Text Key112998296
Report Number3008452825-2018-00224
Device Sequence Number1
Product Code DXO
UDI-Device Identifier05415067025715
UDI-Public05415067025715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/30/2019
Device Model NumberC12059
Device Catalogue NumberC12059
Device Lot Number6240084
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-