Model Number 5392 |
Device Problems
No Display/Image (1183); Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis confirmed the reported issue; there were multiple errors found in the logs.The external pulse generator (epg) failed incoming testing and had an out of specification main printed circuit board (pcb).It was also found that the liquid crystal display (lcd) was damaged and had bleeding pixels.The epg was returned unrepaired at the customer's request.This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator (epg) displayed an error code.The epg has been returned for service.No patient complications have been reported as a result of this event.The epg subsequently tested out of specification during manufacturer's analysis.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|