MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems No Display/Image (1183); Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis confirmed the reported issue; there were multiple errors found in the logs.The external pulse generator (epg) failed incoming testing and had an out of specification main printed circuit board (pcb).It was also found that the liquid crystal display (lcd) was damaged and had bleeding pixels.The epg was returned unrepaired at the customer's request.This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) displayed an error code.The epg has been returned for service.No patient complications have been reported as a result of this event.The epg subsequently tested out of specification during manufacturer's analysis.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7655526
MDR Text Key112995264
Report Number3004593495-2018-00597
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-