• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM Back to Search Results
Catalog Number 12673-03
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/15/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy or inability to maintain position/stability of the device during deployment due to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that an arteriotomy closure of the calcified right common femoral artery was attempted using a proglide device with a 6f sheath after an interventional percutaneous coronary procedure.Reportedly, a cuff miss was noted.Another proglide device was used to achieve hemostasis.There was no reported adverse patient sequela.There was no clinically significant delay in the procedure or therapy.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7655943
MDR Text Key112934070
Report Number2024168-2018-05121
Device Sequence Number1
Product Code MGB
UDI-Device Identifier08717648013089
UDI-Public08717648013089
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number12673-03
Device Lot Number8020141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SHEATH: 6FHEPARIN
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient Weight100
-
-