• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH HEXAPOD STRUT, MEDIUM HOFFMANN LRF 131MM-191MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH HEXAPOD STRUT, MEDIUM HOFFMANN LRF 131MM-191MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 49350030
Device Problems Detachment Of Device Component (1104); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2018
Event Type  Injury  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
Upon a follow-up visit, 3 of 6 hexapod struts were found to have failed.The pins that hold the strut at the hinge were broken.All 6 struts were pulled and replaced.Rep indicated that the pins weren¿t broken.They fell out.They fell completely out of one strut, and halfway out of the other two.
 
Event Description
Upon a follow-up visit, 3 of 6 hexapod struts were found to have failed.The pins that hold the strut at the hinge were broken.All 6 struts were pulled and replaced.Rep indicated that the pins weren¿t broken.They fell out.They fell completely out of one strut, and halfway out of the other two.
 
Manufacturer Narrative
The reported event that hexapod strut, medium hoffmann lrf 131mm-191mm was alleged of 'implant breakage after surgery' could be confirmed.Based on investigation, the root cause was attributed to be manufacturing and supplier related.The failure was caused by manufacturing components out of specifications.The event involved a product problem indicating a non-conformity which is investigated and addressed through a nc and a capa.No further action is required at this time.Product surveillance will continue to monitor complaints of this type for adverse trends.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEXAPOD STRUT, MEDIUM HOFFMANN LRF 131MM-191MM
Type of Device
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7656172
MDR Text Key113003979
Report Number0008031020-2018-00482
Device Sequence Number1
Product Code KTT
UDI-Device Identifier07613327294231
UDI-Public07613327294231
Combination Product (y/n)N
PMA/PMN Number
K163656
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number49350030
Device Lot NumberY08553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Date Manufacturer Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient Weight118
-
-