DEPUY SYNTHES PRODUCTS LLC 11.0 CM LONG ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
Catalog Number LONG |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device is being returned for further investigation.Once the investigation has been completed, a supplemental medwatch will be submitted.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was determined that the nose cone, bearings and extension sleeve of the attachment device were damaged.It was further determined that the balls and cage were missing.It was observed that the ball bearing had fallen apart.It was further determined that the device failed pretest for lock operation and cutter insertion.It was noted in the service order that the device could not be locked.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition regarding the lock was not confirmed.There were no issues observed with the lock operation.Therefore, an assignable root cause was not determined.However, during evaluation, a visual and functional assessment was performed which found that the device had vibration.During service/repair, it was observed that the bearing was corroded and broken.It was determined that the issues were consistent with normal wear.The assignable root cause was determined to be due normal wear out from use.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|