• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107142
Device Problem Misassembled (1398)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/01/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided pictures showed that the effluent line post pump and the pbp line pre pump were inverted.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.The involved operator was retrained and awareness training was provided to all operators.The batch involved in this report was manufactured prior to these actions.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the pbp line and the effluent line were found to be crossed on a prismaflex hf1400 set.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX HF1400
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7656430
MDR Text Key113010311
Report Number8010182-2018-00067
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414069315
UDI-Public(01)07332414069315
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042938
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2020
Device Catalogue Number107142
Device Lot Number18A0501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-