Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Visual inspection noted dried blood/body fluid noted within the lead lumen.The conductor coil was slightly bent approximately 9mm from the lead tip.Foreign material was observed inside the lead lumen near the tip, approximately 14mm from the tip.The foreign material was likely an accumulation of blood/body fluid and polymer coatings separated from the lead and the guide wire, due to the interaction of the wire within the lumen.Laboratory analysis confirmed the obstruction in the lead lumen.
|
Boston scientific received information that this left ventricular (lv) lead was attempted to be implanted.It was noted the patient anatomy was difficult and there was significant scar tissue.The lead was positioned several times due to high pacing thresholds, until an obstruction was noted near the lead tip and a guide wire would no longer pass.The lead was flushed but this did not clear the obstruction.Anew lead was implanted, also with higher threshold values.No adverse patient effects were reported.
|