• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054097
Device Problems Device Alarm System (1012); Incorrect Or Inadequate Test Results (2456); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2018
Event Type  malfunction  
Manufacturer Narrative
Device has been repaired but not returned to the customer, pending customer approval of estimate.
 
Event Description
A ventilator was returned to the manufacturer for service.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, a failure related to the remote alarm connector was observed.The device's interface board was replaced to address the issue.During the evaluation of the device at the manufacturer's service center, the device failed test steps during testing.The device's active exhalation control module was replaced to address the issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin dr.
new kensington, PA 15068
7243349303
MDR Report Key7658537
MDR Text Key113020531
Report Number2518422-2018-01509
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959022751
UDI-Public00606959022751
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054097
Device Catalogue Number1054097
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-