MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 06/06/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the imaging system would intermittently enter stand alone mode.Replacing the umbilical cable resolved the issue.The imaging system then passed the system checkout and was found to be fully functional.The umbilical cable has been received by the manufacturer for evaluation but results are not available yet.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that the system was intermittently losing power and communication.Troubleshooting found that some pins were bent on the umbilical cords.No patient present.
 
Manufacturer Narrative
The analysis of the interface (umbilical) cable was completed by medtronic personnel.Testing found that the cable did not pass bench testing due to four lines exceeding the maximum specified length.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7658960
MDR Text Key113044339
Report Number1723170-2018-03111
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-