AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM Back to Search Results
Catalog Number 12673-05
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy or inability to maintain position/stability of the device during deployment due to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that an arteriotomy closure of the mildly calcified right common femoral artery was attempted using a proglide device via a 7f sheath after an percutaneous transluminal aniogplasty (pta) procedure.Reportedly, a cuff miss occurred.A new proglide device was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7659098
MDR Text Key113041273
Report Number2024168-2018-05131
Device Sequence Number1
Product Code MGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Catalogue Number12673-05
Device Lot Number8042341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SHEATH: 7FHEPARIN
Patient Outcome(s) Required Intervention;
-
-