• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVAILMED 250 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVAILMED 250 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938737
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an exactamix eva (ethylene vinyl acetate) tpn (total parenteral nutrition) bag was leaking from a pinhole on the tubing near the green clamp.The leak was observed after filling with unspecified medication and prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing revealed a leak from a tear/hole at the fill port weld area in the back of the bag.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
250 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
AVAILMED
tijuana, baja california
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
see h10
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7659365
MDR Text Key113256021
Report Number1416980-2018-04046
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477213
UDI-Public(01)00085412477213
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue NumberH938737
Device Lot Number1250830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-