Results: the pet lock was intact on the proximal end of the pusher assembly.The stretch resistance wire (sr wire) was fractured, the embolization coil was detached from its pusher assembly inside its introducer sheath, and the proximal constraint sphere was inside the distal detachment tip (ddt).The introducer sheath was ovalized approximately 6.0 cm from the distal tip.Conclusions: evaluation of the returned ruby coil revealed that the introducer sheath was ovalized.If the device is retracted out of its packaging hoop at extreme angles, damage such as ovalization may occur.Further evaluation of the returned device revealed that the embolization coil sr wire was fractured.If the embolization coil is forcefully retracted against resistance, damage such as this may occur.The embolization coil being retracted through the damage introducer sheath likely contributed to the sr wire fracture.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a coil embolization procedure, the hospital staff found that the introducer sheath of the ruby coil was bent upon removal from the packaging.The damage to the introducer sheath was found prior to use.Therefore, the ruby coil was not used in the procedure.The procedure was completed using a new ruby coil.
|