Model Number TF-25A
Device Problems Microbial Contamination of Device (2303); Torn Material (3024)
Patient Problem Endocarditis (1834)
Event Type  Injury  
Event Description
On (b)(6) 2015, a 25mm trifecta was implanted.In 2018, at an unknown point in time, it is reported the patient underwent redo avr due to a tear along the stent which the user reported was most likely caused by endocarditis with streptococci.The procedure was completed with a non-abbott 25mm magna ease valve.The trifecta valve was sent to hospital bacteriology department for analysis which was inconclusive as the patient had been treated with antibiotics.Additional records were not available.
 
Manufacturer Narrative
Events of severe aortic leakage and a torn leaflet were reported.The results of the investigation are inconclusive since the device was not returned for analysis, however two photos were received for analysis.Based solely on the aforementioned photos, two leaflets appeared to have torn away from a stent post.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7660749
MDR Text Key113152539
Report Number3008452825-2018-00206
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTF-25A
Device Catalogue NumberTF-25A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-