Manufacturer's investigation conclusion: the explanted pump was returned for evaluation.The evaluation of the pump confirmed an issue that could have contributed to the reported pump stoppages on both battery power and the ungrounded cable, which were observed in the submitted system controller log file.The pump was returned assembled with the percutaneous lead (lead) severed approximately 11 inches (") from the pump housing.The remainder of the lead was returned in a segment measuring approximately 29.5¿.The previous percutaneous lead repair was observed approximately 18¿ through 22¿ from the metal connector.Metal braided shield breakdown was observed along the length of the lead with severe breakdown observed adjacent to the nipple of the pump housing, adjacent to the metal connector, and approximately 2.5¿, 18¿, and 27.5¿ from the metal connector.Visual inspection of the underlying wires revealed that the yellow and orange wires were fractured adjacent to the metal connector, and a breach was observed in the insulation of the red wire adjacent to the metal connector, exposing the inner conductors.Hipot testing of the remainder of the percutaneous lead revealed additional breaches in the insulation of the black and green wires adjacent to the nipple of the pump housing, exposing the inner conductors.The observed wire damage appeared consistent with fatigue failure due to repetitive flexing and abrasion against the metal braided shield.If the exposed inner conductors of wires from two different motor phases made direct contact with each other or simultaneous contact with the metal braided shield, the resulting phase-to-phase short could have contributed to the pump stops and low speed hazard/advisory events observed in the submitted log file while connected to any power source.A review of the device history records revealed the device met applicable specifications.No further information was provided.The manufacturer is closing the file on this event.
|