(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Device evaluation summary: the actual device was received for evaluation.A loose needle and a broken suture of product code 1111t, lot# aj9608 were returned for analysis.The loose needle was examined for visual inspection under magnification, the swage and attachment area were as expected.The suture was inspected along the strand, the end of the suture cut (broken) was noted.In addition, several pinch marks near to the thread broken, that appears to be by a surgical instrument could be observed.According to the samples condition, the assignable cause of suture needle pull off is due to improper handling of the sample.When handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments such as forceps or needle holders.
|