Boston scientific received information that there was an alert on this cardiac resynchronization therapy defibrillator (crt-d) related to the right ventricular (rv) lead that is connected to the crt-d.The rv lead is another manufacturer's lead.A code 1004 alert was triggered, as a short circuit was detected.Additionally a code 1007 was triggered indicating that the charge time exceeded 45 seconds, as the charge could not complete.Additional information was reported indicating that on the same day, the battery capacity had depleted, after being implanted for 3 years.The device was operating only in the rv, pacing at 50 paces per minute and therapy was only available in the ventricular fibrillation (vf) zone at 165 beats per minute (bpm).The patient was admitted as they were complaining of chest pain.A few days later, the device failed to detect vf and a shock was externally delivered.The device was subsequently removed and replaced and the leads were surgically abandoned.
|
Upon receipt at our post market quality assurance laboratory, a visual inspection of the device noted no anomalies on the titanium case.Review of the stored memory confirmed a shorted lead indicator and 2 charge time indicators were recorded by the device.A real time x-ray noted that the plasma fuse has been blown open.Based on inspection and analysis of this device, evidence indicates the device was damaged after delivering a shock through a shorted defibrillation lead.The clinical observations occurred as a result of the shorted lead.The lead is from another manufacturer.
|