MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAL
Device Problems Device Inoperable (1663); Battery Problem (2885)
Patient Problem Hyperglycemia (1905)
Event Date 06/06/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had a keypad anomaly.The buttons were not responsive.The insulin pump alarmed battery out limit.Customer's blood glucose level was 600 mg/dl.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
The information provided was incorrect with the initial report.The correct information has been provided with this report.Insulin pump was received with no button response at act and up arrow button due to unlocked connector on lcd board.No button error alarm during testing.Unable to confirm unexpected battery out limit alarm and unable to perform any tests due to button unresponsive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7663263
MDR Text Key113160086
Report Number3004209178-2018-88340
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169507234
UDI-Public(01)00643169507234
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAL
Device Catalogue NumberMMT-723NAL
Device Lot NumberB5723NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
Patient Weight89
-
-