Model Number 0165SI16 |
Device Problems
Inflation Problem (1310); Failure to Infuse (2340); No Flow (2991)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that water would not inject into the balloon, and the balloon could not be inflated after catheter placement.
|
|
Event Description
|
It was reported that water would not inject into the balloon, and the balloon could not be inflated after catheter placement.
|
|
Manufacturer Narrative
|
Received only catheter with smear on cap for evaluation.The evaluation found that the catheter could be inflated without difficulty.The catheter was dissected and found no conditions that would contribute to the reported problem.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "(6) insert catheter into urethral meatus, and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a syringe packaged in the tray, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.".
|
|