• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI16
Device Problems Inflation Problem (1310); Failure to Infuse (2340); No Flow (2991)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that water would not inject into the balloon, and the balloon could not be inflated after catheter placement.
 
Event Description
It was reported that water would not inject into the balloon, and the balloon could not be inflated after catheter placement.
 
Manufacturer Narrative
Received only catheter with smear on cap for evaluation.The evaluation found that the catheter could be inflated without difficulty.The catheter was dissected and found no conditions that would contribute to the reported problem.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "(6) insert catheter into urethral meatus, and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a syringe packaged in the tray, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7663614
MDR Text Key113454871
Report Number1018233-2018-02591
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016752
UDI-Public(01)00801741016752
Combination Product (y/n)N
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 07/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0165SI16
Device Catalogue Number0165SI16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Date Manufacturer Received07/17/2018
Is This a Reprocessed and Reused Single-Use Device? No