The customer contacted the siemens customer care center (ccc).The ccc specialist was made aware that the aliquotter belt was broken and replaced, and that laboratory personnel did not have to evacuate as a result of the spark and smoke that was emitted by the instrument.A siemens customer service engineer (cse) was dispatched to the customer site.The cse inspected the instrument and resolved the issue by replacing the control servo hybrid printed circuit board, chain cable timing 4p x 2500l poly, c-drive vertical assy aliquotter, ionizing fan service kit, aligning the auto and manual aliquot probe and the reagent shuttle 2.The issue is resolved, and the instrument is operational.No further evaluation of this device is required.
|