It was reported that low output current was seen during implant of the generator.Data showed that no diagnostics were performed during the implant, so it could not be determined if the output current was delivered properly after the initial interrogation.Lead impedance was within normal limits, indicating that the lead was functioning properly.No further relevant information has been received to date.
|
Data from the surgery was reviewed, and diagnostics were not performed on the generator.Therefore, the impedance value was only a stored value from manufacturing and not representative of the actual system impedance.Also, the output current delivered upon initial interrogation was 0ma, but the generator should have tried to deliver two on times of stimulation without success.The next interrogation had the same results of 0ma delivered.However, a final interrogation immediately after showed that the device delivered 2ma successfully.Without diagnostic results, it cannot be determined what caused the device to be unable to deliver the stimulation initially.No further relevant information has been received to date.
|