• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-M
Device Problems Defective Component (2292); Difficult to Open or Close (2921); Mechanical Jam (2983)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the physician attempted to use a h1-m to treat a lesion in the superficial femoral artery(sfa) as per the ifu.No abnormalities were reported in relation to the patient anatomy.It was reported the thumbswitch on the hawkone device jammed while in use in the patient and would not fully close or open.The device was removed safely from the patient.The device was cleaned but remained stuck.The thumbswitch would not push all the way forward or back, so tissue could not be completely flushed.The customer verified the nosecone was not full.No deformation was noted to the cutter and no pieces of the cutter were reported to be missing the procedure was completed prior to the hawkone thumbswitch malfunction.It was post-dilated with pta.No cine images are available.No patient injury was reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7669077
MDR Text Key113636405
Report Number9612164-2018-01658
Device Sequence Number1
Product Code MCW
UDI-Device Identifier00643169968332
UDI-Public00643169968332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/09/2021
Device Catalogue NumberH1-M
Device Lot Number0009076185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Date Device Manufactured04/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-