MAUDE Adverse Event Report: ARJO HOSPITAL EQUIPMENT AB PARKER BATH; BATH, HYDRO-MASSAGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJO HOSPITAL EQUIPMENT AB PARKER BATH; BATH, HYDRO-MASSAGE Back to Search Results
Model Number AL14010-US
Device Problems Mechanical Problem (1384); Migration (4003)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The review of similar reportable events with the involvement of the parker bath in last years revealed a low number of complaints related to the situation where the defective door gas strut contributed to an event occurrence.The malfunction was detected during servicing of the device by arjo representative.As a correction, the worn gas strut was replaced.According to arjo technician opinion, the faulty gas strut had low resistance.The faulty part was disposed after the repair.The previous bathtub preventive maintenance was performed by arjo representative in (b)(6) 2017.This device was also repaired in (b)(6) 2018, when the drainage hose was replaced because of the damage.According to the service summary no malfunction of the door gas strut was indicated in both cases.In conclusion, the review of service history for the bathtub did not show any replacements of the gas strut in the past.The parker bath is subject to wear and tear, and recommended maintenance instructions must be performed when specified to ensure that the equipment remains within its original manufacturing specification.List of recommended steps can be found in product preventive maintenance schedule active at the time when this device was manufactured (pm0684_01 dated on december 2002).In order to ensure the efficiency and functionality of the door strut, this particular component should be checked every 12 months according the following instruction: "open the door and check that the gas strut holds the door in the upright position and does not drop." it informs also that the door strut should be replaced after 48 months/4 years of functioning.Please note that operating and daily maintenance instructions (ifu, 04.Al.00/1 dated on june 2003) delivered with the device, include information which each user of the arjo equipment should follow.In connection with the subject of this investigation, the following warnings were established to prevent from any injury occurrence (p.4): "always ensure that the equipment is handled by trained staff." "always ensure that the bathers limbs are clear of the door before closing." "always keep fingers clear of the door when closing." the preventive maintenance schedule also reminds the customer to check the operation of the door regularly on a weekly basis to detect any failure related to this assembly: "open and close the door and check the lock and supporting gas strut for correct and smooth operation." following the collected information it can be determined that the preventive maintenance of the involved unit was not correctly performed.Therefore the most likely root cause for this case is related to maintenance deficiencies.At the time of event the device did not meet manufacturer specification due to worn door gas strut.This however did not cause or contribute to adverse event.According to the collected information the bathtub was used for patient hygiene prior to malfunction detection.We report this incident in abundance of caution due to malfunction which could pose a risk for device users.
 
Event Description
Arjo received a customer complaint on parker bath, where it was reported that the tub door was falling (not keeping the upright position) due to worn gas strut.Despite this malfunction, the bathtub was still used by the customer facility.No event was experienced by the device's users, no harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARKER BATH
Type of Device
BATH, HYDRO-MASSAGE
Manufacturer (Section D)
ARJO HOSPITAL EQUIPMENT AB
verkstadsvagen 5
eslov, 24121
SW  24121
Manufacturer (Section G)
ARJO HOSPITAL EQUIPMENT AB
verkstadsvagen 5
eslov, 24121
SW   24121
Manufacturer Contact
kinga stolinska
ks. wawrzyniaka 2
komorniki, 62-05-2
PL   62-052
MDR Report Key7669162
MDR Text Key113905094
Report Number3007420694-2018-00137
Device Sequence Number1
Product Code ILJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberAL14010-US
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/09/2018
Distributor Facility Aware Date06/18/2018
Device Age13 YR
Event Location Nursing Home
Date Report to Manufacturer07/09/2018
Date Manufacturer Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-