TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801764 |
Device Problems
Failure to Run on Battery (1466); Connection Problem (2900)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the battery icon was red.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit was not switching over to battery when there was loss of power.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
As per manufacturer's subsidiary, there was a loose connection on the power manager board.The connection was tightened and the unit operated as intended.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|