ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3228
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hematoma (1884); Paralysis (1997); Weakness (2145)
Event Date 06/14/2018
Event Type  Injury  
Manufacturer Narrative
Manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient experienced paralysis in both legs as well as weakness and a hematoma at t8 lead site (reference mfr.Report# 1627487-2018-06359).The physician removed the hematoma and repositioned the lead in the original location.Intra-op testing revealed leads were within normal range.However, paralysis still reside post revision.
 
Event Description
Additional information received identified the issue was resolved post evacuation of hematoma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
falon whitfield
6901 preston road
plano, TX 75024
9725268233
MDR Report Key7674681
MDR Text Key113569073
Report Number1627487-2018-06358
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017253
UDI-Public05415067017253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/21/2020
Device Model Number3228
Device Lot Number6324988
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3662, SCS IPG
Patient Outcome(s) Hospitalization; Life Threatening; Other;
Patient Age45 YR
-
-