Model Number W4TR04 |
Device Problems
Electrical /Electronic Property Problem (1198); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy pacemaker (crt-p) experienced a partial electrical reset.The reset was cleared and the crt-p remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a software error.Analysis of the device memory indicated that the atrial and ventricular rate histogram data were missing/invalid.Analysis of the device memory indicated an issue with missing/invalid diagnostic data.This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|