The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the spring in the red cpc connector on the tah-t was sideways.The customer also reported that the cpc connector was replaced.There was no reported adverse patient impact.
|
Visual inspection confirmed the customer-reported sideways spring, as the spring inside the female cpc connector was displaced sideways inside the housing.Clinicians and patients are trained to thread a wire tie through the thumb release tab of the cpc connectors to prevent inadvertent disconnection of the cannulae from the drivelines.It is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|