SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 500101-001
Device Problems Mechanical Problem (1384); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/18/2018
Event Type  malfunction  
Manufacturer Narrative
The cpc connector has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow up mdr.(b)(4) initial.
 
Event Description
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the spring in the red cpc connector on the tah-t was sideways.The customer also reported that the cpc connector was replaced.There was no reported adverse patient impact.
 
Manufacturer Narrative
Visual inspection confirmed the customer-reported sideways spring, as the spring inside the female cpc connector was displaced sideways inside the housing.Clinicians and patients are trained to thread a wire tie through the thumb release tab of the cpc connectors to prevent inadvertent disconnection of the cannulae from the drivelines.It is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7680239
MDR Text Key114025297
Report Number3003761017-2018-00400
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number107855
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Date Manufacturer Received06/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age48 YR
-
-