• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI16
Device Problems Deflation Problem (1149); No Flow (2991)
Patient Problems Patient Problem/Medical Problem (2688); No Known Impact Or Consequence To Patient (2692)
Event Date 06/19/2018
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the balloon did not deflate on the 17th day of placement.Initially only 2cc of water was aspirated.The inflation funnel was cut and a guidewire was inserted in the lumen, however no water released.Two days later, the balloon was punctured under ultrasound guidance and the catheter was removed.
 
Event Description
It was reported that the balloon did not deflate on the 17th day of placement.Initially only 2cc of water was aspirated.The inflation funnel was cut and a guidewire was inserted in the lumen, however no water released.Two days later, the balloon was punctured under ultrasound guidance and the catheter was removed.
 
Manufacturer Narrative
The reported event was found inconclusive due to the way the sample was received.The sample was evaluated and no pinch or blockage observed.High salt accumulation was noted inside of the drainage eye, drainage lumen and around the balloon surface.Water was able to be introduce in all returned pieces.No conditions found on sample that could be associated with reported event.Due to poor condition of the returned sample, a complete evaluation could not be performed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "2.Precautions for use (1) when resistance is encountered in inserting catheter, stop the procedure and remove the catheter.(2) when deflating balloon, do not aspirate with a syringe.[the inflation lumen for balloon deflation may be occluded by negative pressure, and as a result the catheter cannot be removed.] (3) no substance except sterile water should be used to inflate the balloon.[if contrast medium is used, balloon may burst.If normal saline is used, crystallized salt may occlude the inflation lumen to prevent deflation of the balloon.If air is used, air may escape to cause inadvertent deflation of the balloon so that the catheter may come out prematurely.] (4) do not wipe catheter surface with organic solvents such as alcohol.(5) do not aspirate urine through drainage funnel wall." the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7680750
MDR Text Key113742043
Report Number1018233-2018-02741
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016752
UDI-Public(01)00801741016752
Combination Product (y/n)N
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2021
Device Model Number0165SI16
Device Catalogue Number0165SI16
Device Lot Number7JW7127
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Date Manufacturer Received07/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-