Model Number 383401 |
Device Problem
No Display/Image (1183)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The pump was returned back to stock with no specific error reported.Upon triage on (b)(6) 2018, service found the pump¿s screen was blank.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo joey pump was performed for the reported condition of ¿service found the pump¿s screen was blank.¿ the unit was triaged and the reported condition was confirmed.Unit was powered up on battery and no image was shown on the display.A component in the vdc circuit was showing slight discoloration from overheating and was hot to the touch when the unit was on.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|