It was reported that a male patient in the age range between 25 to 35 years old with mental disorder had a broken implant after a suicide attempt.The patient arrived to the hospital with several fractures, and after x-rays examination, the implant was found to be broken and a surgery was performed to explant the unit.Chronology: on (b)(6) 2018 a radial head exchange surgery was performed with the implant.On (b)(6) 2018 during routine check, the implant head was found to be "discharged".On (b)(6) 2018 additional surgery was performed in order to remove the unit and place a new implant.No injury to patient reported and his condition is good.
|
The manufacturing records for lot number 221675 was reviewed.A quantity of 2 devices were inspected on 06/05/2015.No abnormalities or nonconformances were identified that would have caused or contributed to the event.The implant was assessed by engineering and no out of tolerance conditions, defects, abnormalities or damage could be identified.The most likely failure mode here was loosening, rather than implant breakage, seeing as there is no evidence of breakage on the returned parts.Based on the assessment performed on the returned device , integra is unable to confirm the reported non-conformance or determine root cause.Base upon the assessment of the returned the device, the root cause was most likely loosing or dissociation possibly from surgical technique, patient after care compliance, or patient anatomy.
|