• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Death (1802)
Event Date 06/19/2018
Event Type  Death  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 3 of 3.Reference mfr.Report#1627487-2018-06259, reference mfr report#3006705815-2018-01589.It was reported by the patient¿s family member that the patient passed away on (b)(6) 2018.The cause of death is unknown at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
seema wakdikar
9725268205
MDR Report Key7685922
MDR Text Key113910929
Report Number3006705815-2018-01590
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2018
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Expiration Date11/13/2019
Device Model Number3186
Device Lot NumberA000047732
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192(2)SCS ANCHOR
Patient Outcome(s) Death;
-
-