• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 295251
Device Problems Break (1069); Material Split, Cut or Torn (4008)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reports that the unit has a damaged power cord with copper exposed.
 
Manufacturer Narrative
An evaluation of the scd 700 compression system was performed for the reported condition of, ¿the unit has a damaged power cord with copper exposed¿.The unit was triaged and service confirmed the reported issue.The power cord was replaced.Device fully checked for functionality and patient electrical safety according to manufacturer's guideline.Power cords periodically require replacement due to age, usage and user damage.Kinking of the cable assembly along the cord length, which is indicative of being tightly wrapped around the bed hook has been identified as a contributing factor for this failure mode.Over time, tightly wrapping of the cord around the bed hook causes a constant tension to be applied to the outer cover resulting in the cover being milked out of the strain relief.Stretching the power cord at a sharp angle to the plug to reach the wall outlet has the same effect.Therefore the potential root cause for this issue is customer misuse.A review of the device history record shows that this unit was manufactured and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Type of Device
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7689614
MDR Text Key114285212
Report Number3006451981-2018-00507
Device Sequence Number1
Product Code JOW
UDI-Device Identifier10884521129580
UDI-Public10884521129580
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number295251
Device Catalogue Number295251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2018
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-