Model Number BL15455 |
Device Problems
Use of Incorrect Control/Treatment Settings (1126); Use of Device Problem (1670)
|
Patient Problem
Capsular Bag Tear (2639)
|
Event Date 06/13/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
The system was inspected at the site by a bausch + lomb representative.The system was found to function within specification with no faults found.A review of the manufacturing records found the system met all specification at the time of release.It was determined the machine settings used during the procedure were incorrect.The surgeon has been advised of the correct machine settings.
|
|
Event Description
|
A report received from a user facility in the (b)(4) stated that according to the surgeon, the phaco phase in the stellaris elite foot control settings is activated too early which caused the surgeon to inadvertently phaco the capsular bag.Additional information regarding patient outcome has been requested.
|
|
Manufacturer Narrative
|
We have contacted the hospital several times for an update on the patient's conditions; however, no updates were provided.A follow-up visit was set up by our territory manager to verify the unit performance.The review indicated that the problem was associated to the incorrect surgeon setting for the foot control with serial number (b)(4).
|
|
Search Alerts/Recalls
|