Model Number 5076-58 |
Device Problems
Device Alarm System (1012); High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: unknown biotronik lead, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that prior to implant, the right ventricular (rv) lead triggered an alerted for bipolar lead impedance measurement that was undefined over 3000 ohms.Presently, the lead measurements appear to be within expected range.The lead remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|