Model Number 5392 |
Device Problems
Loose or Intermittent Connection (1371); Structural Problem (2506); Material Protrusion/Extrusion (2979)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: it was confirmed that the case seal was protruding.Plug and case screws were contaminated.Lower case bosses were broken.The upper case was cracked.Display wire was pinched.All found defective parts were replaced and all other identified issues were resolved.The device was re-calibrated and functionally tested and it passed its final quality assurance tests.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator's (epg) case seal was protruding.The programmer was returned for service.No patient complications have been reported as a result of this event.The programmer tested out of specification during manufacturer's analysis.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|