Model Number 179706990
Device Problems Break (1069); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination of the viper screw reveals signs of operative use as evidenced by superficial markings, with its threads completely peeled off and damage to the drive feature of the screw head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn threads on the viper screw cannot be determined.This damage may have potentially occurred due to cross-threading a set screw upon insertion.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that while the surgeon was torqueing the set screw at the end of the case at normal 80 then wanted to torque at 100.When doing so, the set screw stripped out and compromised the inner screw threads of the screw.Pulled the screw out and replaced it.No adverse event and prolonged surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER T27 ILIAC SCREW 9X90, TI
Type of Device
ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS SARL INT'L
chemin blanc 38
le locle
SZ  
Manufacturer Contact
jason busch
325 paramount drive
raynham, MA 02767
5088808100
MDR Report Key7695385
MDR Text Key114678032
Report Number1526439-2018-50696
Device Sequence Number1
Product Code MNH
UDI-Device Identifier10705034266589
UDI-Public(01)10705034266589
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Caregivers
Type of Report Initial
Report Date 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179706990
Device Catalogue Number179706990
Device Lot NumberTBTKX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/28/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-