MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX M 100BX; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX M 100BX; MALE EXTERNAL CATHETER Back to Search Results
Model Number 33102
Device Problems Nonstandard Device (1420); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the adhesive on the male external catheter was strong.
 
Manufacturer Narrative
The reported event was unconfirmed.Visual evaluation of the samples noted no obvious defects.Adhesive peel test was performed and the samples were tested according to ansi/asq z1.4-2003 (general inspection level ii/single sampling plan).Samples were cut (0.7090", 0.6995") to the required width (0.70" +/- 0.05 ").Adhesive peel strength (0.6834 lbf, 0.7684 lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: wash penis with mild soap and warmwater.Dry thoroughly.Trim pubic hair if necessary.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Connect to drainage device.Directions: to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: additional event information.
 
Event Description
It was reported that the adhesive on the male external catheter was strong.
 
Event Description
It was reported that the adhesive on the male external catheter was strong.
 
Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the adhesive on the male external catheter was strong.
 
Manufacturer Narrative
The reported event was unconfirmed.Visual evaluation of the samples noted no obvious defects.Adhesive peel test was performed and the samples were tested according to ansi/asq z1.4-2003 (general inspection level ii/single sampling plan).Samples were cut (0.7090", 0.6995") to the required width (1.00" +/- 0.05 ").Adhesive peel strength (0.6834 lbf, 0.7684 lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: concomitant medical products and device evaluated by mfr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC ULTRAFLEX M 100BX
Type of Device
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7697971
MDR Text Key114675972
Report Number1018233-2018-02864
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741070679
UDI-Public(01)00801741070679
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup,Followup
Report Date 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2023
Device Model Number33102
Device Catalogue Number33102
Device Lot NumberJUCQ0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Date Manufacturer Received08/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-