Model Number 33102 |
Device Problems
Nonstandard Device (1420); Material Integrity Problem (2978)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that the adhesive on the male external catheter was strong.
|
|
Manufacturer Narrative
|
The reported event was unconfirmed.Visual evaluation of the samples noted no obvious defects.Adhesive peel test was performed and the samples were tested according to ansi/asq z1.4-2003 (general inspection level ii/single sampling plan).Samples were cut (0.7090", 0.6995") to the required width (0.70" +/- 0.05 ").Adhesive peel strength (0.6834 lbf, 0.7684 lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: wash penis with mild soap and warmwater.Dry thoroughly.Trim pubic hair if necessary.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Connect to drainage device.Directions: to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: additional event information.
|
|
Event Description
|
It was reported that the adhesive on the male external catheter was strong.
|
|
Event Description
|
It was reported that the adhesive on the male external catheter was strong.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that the adhesive on the male external catheter was strong.
|
|
Manufacturer Narrative
|
The reported event was unconfirmed.Visual evaluation of the samples noted no obvious defects.Adhesive peel test was performed and the samples were tested according to ansi/asq z1.4-2003 (general inspection level ii/single sampling plan).Samples were cut (0.7090", 0.6995") to the required width (1.00" +/- 0.05 ").Adhesive peel strength (0.6834 lbf, 0.7684 lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: concomitant medical products and device evaluated by mfr.
|
|