• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION FREESTYLE AORTIC ROOT BIOPROSTHESIS; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION FREESTYLE AORTIC ROOT BIOPROSTHESIS; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number FR995-27
Device Problems Structural Problem (2506); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Thrombus (2101); Tissue Damage (2104); Pseudoaneurysm (2605)
Event Date 06/13/2018
Event Type  Injury  
Manufacturer Narrative
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that 10 years and 5 months post implant of this aortic bioprosthetic valve, the valve was explanted and replaced with a bioprosthetic valve of the same model.The physician reported the replacement was due to a false aneurysm caused by wall separation of the right coronary ostium. old and recent thrombus was noted at the false aneurysm.It was also reported that the right coronary ostium was completely separated from the bioprosthetic root, the entire right sinus of valsalva was missing, and attempts to mobilize the inflamed left coronary ostium failed.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE AORTIC ROOT BIOPROSTHESIS
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key7698606
MDR Text Key114323617
Report Number2025587-2018-01780
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00681490265942
UDI-Public00681490265942
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/17/2012
Device Model NumberFR995-27
Device Catalogue NumberFR995-27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Date Device Manufactured11/20/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient Weight80
-
-