MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problems
Break (1069); Device Slipped (1584); Failure to Advance (2524); Material Deformation (2976); Material Integrity Problem (2978)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/25/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that prior to a cryo ablation procedure, there was a slight kink on the mapping catheter.Additionally, the tip of the catheter came out from the inserter, and upon retracting it into the inserted, it became stuck.There were several attempts made to advance the catheter tip out from the inserter, and it was observed the tip of the catheter was broken.The mapping catheter wasreplaced.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device and data files were returned and analyzed.Data files showed that 7 applications were performed with a balloon catheter on the date of the event with no issues.Visual inspection of the mapping catheter showed that there was no kink on the shaft and no introducer issue.Also, the mapping catheter was kinked on the loop/misshape and the pebax tube was ribbed.The loop tip was damaged, detached and the electrode was crushed.In conclusion, the mapping catheter failed the return test due to a kink on the loop, a damaged pebax, detached/ damaged loop and a crushed electrode.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|