Model Number 4068 |
Device Problems
Signal Artifact/Noise (1036); Electrical /Electronic Property Problem (1198); Over-Sensing (1438); Low impedance (2285); Device Displays Incorrect Message (2591); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the right atrial (ra) lead exhibited oversensing with noise.It was also noted that the polarity switched to unipolar and there were lots of low impedance warnings.The lead remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated atrial oversensing.Analysis of the device memory indicated the impedance of the atrial pacing lead was below the expected lower range.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|