• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Catalog Number BI70000027230
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 06/27/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the door winch assembly was replaced to restore functionality.The imaging system then passed the system checkout and was found to be fully functional.The component has not been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information that, while outside of a procedure, the door winch assembly of the imaging system was damaged.There was no patient present when this issue was identified.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7705028
MDR Text Key114543279
Report Number1723170-2018-03605
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBI70000027230
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-